Investigation of scatter from out of the field of view and multiple scatter in PET using Monte Carlo simulations

Get Citation
YE Ting, CHAI Pei, GAO Juan, YUN Ming-Kai, LIU Shuang-Quan, SHAN Bao-Ci and WEI Long. Investigation of scatter from out of the field of view and multiple scatter in PET using Monte Carlo simulations[J]. Chinese Physics C, 2011, 35(12): 1166-1171. doi: 10.1088/1674-1137/35/12/017
YE Ting, CHAI Pei, GAO Juan, YUN Ming-Kai, LIU Shuang-Quan, SHAN Bao-Ci and WEI Long. Investigation of scatter from out of the field of view and multiple scatter in PET using Monte Carlo simulations[J]. Chinese Physics C, 2011, 35(12): 1166-1171.  doi: 10.1088/1674-1137/35/12/017 shu
Milestone
Received: 2011-02-14
Revised: 2011-03-08
Article Metric

Article Views(2308)
PDF Downloads(382)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Investigation of scatter from out of the field of view and multiple scatter in PET using Monte Carlo simulations

Abstract: In fully three-dimensional (3D) positron emission tomography (PET) imaging, the scatter fraction (SF) is about 40%-60%, which may degrade the imaging quality severely. Scatter correction is important for high quality image reconstruction. Model-based scatter correction has been proved to be accurate and available in clinical PET. However, it does not correct the scatter from out of the field of view (OFOV) and multiple scatters. In this study, we demonstrate the radial and axial distribution of scatters from OFOV when the source is located in different radial positions. In order to apply the above conclusions to different PET systems, we characterize the scatters from OFOV as a function of the ratio of the scanner diameter to the length of the axial field of view (AFOV) by modeling several typical whole-body and micro PET systems. The proportions of true events (S0-0), single scatter of one photon (S1-0) , single scatter of both photons (S1-1) , double scatter of one photon (S2-0) and multiple scatter (Sm) are also calculated and compared. Here the 3D-PET Monte Carlo simulations are performed with the Geant4 Application for Tomography Emission (GATE). In summary, the scatters from OFOV tend to be recorded on the lines of response (LOR) far away from the source. They have a much more serious impact on whole-body PET than micro PET depending on the ratio of scanner diameter to the length of AFOV. In whole-body PET, twice scatters including single scatter of both photons (S1-1) and double scatter of one photon (S2-0) add up to about 12% so that twice scatter correction must be taken into account to acquire a high quality reconstruction image.