Performance of a 3 mm×3 mm silicon photomultiplier for use on the X-ray calibration system of the SVOM gamma ray monitor
- Received Date: 2011-06-29
- Accepted Date: 2011-07-25
- Available Online: 2012-04-05
Abstract: The calibration detector of a gamma ray monitor (GRM) is designed to detect alpha particles from 241Am and to send out the coincidence signal to the GRM X-ray detector. The silicon photomultiplier (SiPM), as a novel photon device, is a good candidate to convert alpha-exciting fluorescent photons into electric signals. Three types of SiPMs from SSPM and MPPC, each having an active area of 3 mm×3 mm, were compared in the matter of the spectra from low-intensity light, dark count, crosstalk probability and I-V curve. The temperature coe cient of SSPM-0710G9MM was also characterized. The application of a SiPM on the GRM has been proved to be feasible.