Study on gamma response function of EJ301 organic liquid scintillator with GEANT4 and FLUKA

Get Citation
ZHANG Su-Ya-La-Tu, CHEN Zhi-Qiang, HAN Rui, LIU Xing-Quan, LIN Wei-Ping, JIN Zeng-Xue, XI Yin-Yin, LIU Jian-Li and SHI Fu-Dong. Study on gamma response function of EJ301 organic liquid scintillator with GEANT4 and FLUKA[J]. Chinese Physics C, 2013, 37(12): 126003. doi: 10.1088/1674-1137/37/12/126003
ZHANG Su-Ya-La-Tu, CHEN Zhi-Qiang, HAN Rui, LIU Xing-Quan, LIN Wei-Ping, JIN Zeng-Xue, XI Yin-Yin, LIU Jian-Li and SHI Fu-Dong. Study on gamma response function of EJ301 organic liquid scintillator with GEANT4 and FLUKA[J]. Chinese Physics C, 2013, 37(12): 126003.  doi: 10.1088/1674-1137/37/12/126003 shu
Milestone
Received: 2013-03-01
Revised: 1900-01-01
Article Metric

Article Views(2248)
PDF Downloads(639)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Study on gamma response function of EJ301 organic liquid scintillator with GEANT4 and FLUKA

    Corresponding author: CHEN Zhi-Qiang,

Abstract: The gamma response function is required for energy calibration of EJ301 (5 cm in diameter and 20 cm in height) organic liquid scintillator detector by means of gamma sources. The GEANT4 and FLUKA Monte Carlo simulation packages were used to simulate the response function of the detector for standard 22Na, 60Co, 137Cs gamma sources. The simulated results showed a good agreement with experimental data by incorporating the energy resolution function to simulation codes. The energy resolution and the position of the maximum Compton electron energy were obtained by comparing measured light output distribution with simulated one. The energy resolution of the detector varied from 21.2% to 12.4% for electrons in the energy region from 0.341 MeV to 1.12 MeV. The accurate position of the maximum Compton electron energy was determined at the position 81% of maximum height of Compton edges distribution. In addition, the relation of the electron energy calibration and the effective neutron detection thresholds were described in detail. The present results indicated that both packages were suited for studying the gamma response function of EJ301 detector.