-
The experiment was carried out at the HI-13 tandem accelerator facility at the China Institute of Atomic Energy, Beijing. A
$ ^{16} $ O beam at an energy of 96 MeV and with an intensity of about 4 enA was impinged onto a self-supporting$ ^6 $ LiF target (275 μg/cm$ ^2 $ ). A$ ^6 $ Li($ ^{16} $ O,$ ^{19} $ Ne$ ^* $ )$ ^3 $ H transfer reaction was conducted, followed by cluster decay from the$ ^{19} $ Ne resonant states.As shown in Fig. 1, a total of eight position-sensitive charged-particle telescopes (L0-L3 and R0-R3 detector arrays; L and R denote the left and right sides of the beam line, respectively) were symmetrically installed on both sides of the beam line. The two forward telescopes (L0 and R0), located at a distance of 160 mm from the target and centered at
$ \pm 24° $ with respect to the beam line, were composed of two layers of double-sided silicon-strip detectors (DSSDs) with thicknesses of about 40 μm (W1-40) and 500 μm (BB7-500), respectively, and one large size silicon detector (SSD) with a thickness of 1500 μm (MSX40-1500). The active areas are$ 50\; \rm{mm} \times 50 \; \rm{mm} $ for W1-40 and$ 64\; \rm{mm} \times 64\; \rm{mm} $ for BB7-500 and MSX40-1500. Each side of W1-40 (BB7-500) is divided into 16 (32) strips, providing good two-dimensional position resolution. The L1 and R1 telescopes were located at 180 mm from the target and centered at$ \pm 47° $ , and the detector composition was the same as that for L0 and R0. The L2 and R2 telescopes, each composed of a W1-40 and a MSX40-1500, were located at 180 mm from the target and centered at$ \pm 70 ° $ . The L3 and R3 telescopes, each composed of a BB7-40 and a MSX40-1500, were located at 180 mm from the target and centered at$ \pm 110 ° $ .The
$ ^3 $ H particles, recoiled from the$ ^6 $ Li($ ^{16} $ O,$ ^{19} $ Ne$ ^* $ )$ ^3 $ H transfer reaction, can be detected by the large angle telescopes (L1-3 and R1-3). The α particle or proton, emitted from the$ ^{19} $ Ne-resonance decay ($ ^{19}\rm{Ne} \to \alpha + ^{15}\rm{O} $ and$ ^{19}{\rm Ne} \to p + ^{18}\rm{F} $ ), were mostly measured by the forward telescopes L0 and R0. According to the reaction kinematics, all these light particles could penetrate the first layer but stopped in the last layers of the telescopes and, therefore, can be well identified through the standard energy loss versus residual energy ($ \Delta E - E $ ) method [19−24]. The particle identification (PID) performances of the L0 and L1 telescopes are shown in Fig. 2 (a) and (b), respectively.Figure 2. (color online) PID spectra measured by (a) the first two layers of L0 telescope and (b) the last two layers of L1 telescope.
Using the information of the detected recoil particle
$ ^3 $ H and decay particle α (or p), together with the beam information, the momentum and energy of the remaining decay fragment$ ^{15} $ O (or$ ^{18} $ F) can be deduced according to the momentum conservation law [20, 25]. The reaction energy, namely the Q value, can then be calculated by subtracting the beam energy from the total kinematic energy of the final three particles [19−23, 26]:$ \begin{equation} Q=E_{\rm tot}-E_{\rm beam}=\sum_{i} E_i - E_{\rm beam}, \end{equation} $
(1) where
$E_{\rm beam}$ and$ E_i $ ($ i = 1\; \rm{to}\; 3 $ ) are the kinetic energies of the incident beam particle and the final three particles. Figure 3 shows the Q-value spectra for the two reaction-decay channels, respectively, as exhibited by the corresponding peaks. In the spectra, we see also quite intense and broadly distributed background beneath the expected Q-value peaks. This should be attributed to the misidentified particles in the PID spectra and the undetected particles that do not belong to the assumed decay channels. By gating on the specific Q-value peak, the mass combination of the reaction channel can be selected [19, 20, 24]. The Q-value analysis also demonstrated the negligible contribution from the$ ^{19} $ F content of the target material to the present measurement, which was also verified by the EP-plot analysis [21, 27]. -
The excitation energy spectra of
$ ^{19} $ Ne in the α decay channel as well as the proton decay channel can be obtained by the standard invariant mass (IM) method [19−24, 26, 28−30]. The invariant mass M of the residual nucleus, the relative energy of its decay fragments ($ E_{\rm{rel}} $ ) and the excitation energy ($ E_{x} $ ) of the mother nucleus (residual nucleus) are defined according to [24, 26, 28−31]$ \begin{align} &M^2=(\sum_i E_i)^2-(\sum_i \boldsymbol{p}_i)^2, \end{align} $
(2) $ \begin{align} &E_{\rm{rel}}=M-\sum_i m_i ,\end{align} $
(3) $ \begin{align} &E_{\rm{x}}=E_{\rm{rel}}+E_{\rm{thres}}, \end{align} $
(4) where
$ E_i $ ,$ p_i $ , and$ m_i $ (i = 1, 2) are the kinetic energy, momentum, and mass, respectively, of the decay fragment, and$ E_{\rm{thres}} $ is the threshold energy for separating the residual nucleus into the corresponding two fragments.Figure 4 (a) and (b) represent the excitation energy spectra of
$ ^{19} $ Ne obtained from the IM reconstruction using the detected α particle plus the deduced$ ^{15} $ O fragment and that obtained using the detected proton plus the deduced$ ^{18} $ F fragment, respectively [20]. Each spectrum is fitted with several peak functions plus a smoothly varying background curve [32]. Each peak function is a convolution of the Breit-Wigner (BW) form and the energy-resolution Gaussian function [20, 25, 32]. The initial values for the peak central positions were set according to the previously reported results in the literature or the actual spectrum shape. The widths (Γ) of the BW forms were fixed according to values available from the literature (Table 1) or otherwise used as free fitting parameters. The experimental energy resolutions, as well as the detection efficiencies (black-dotted lines in Fig. 4), as a function of the relative energy$ E_{\rm{rel}} $ were obtained by Monte Carlo simulation, taking into account the realistic detector setup and performances. The obtained peak positions are listed in Table 1. The statistical errors for the peak positions obtained from the fitting are very small in comparison to the typical systematic uncertainty. The latter is estimated to be about 100 keV, and the main contribution is from the energy calibration of the silicon detectors [19]. Since the widths of the observed peaks are essentially dominated by the resolution function, which is about 300−350 keV (FWHM) in the considered energy range, the Γ parameters extracted from the fitting procedure are mostly very small and with large uncertainties. We therefore do not list them in Table 1.Figure 4. (color online) Excitation energy spectra of
$^{19}$ Ne based on the reconstructions from (a) the detected α particle and the deduced$^{15}$ O fragment; (b) the detected proton and the deduced$^{18}$ F fragment. Each spectrum is fitted by the sum (red-solid line) of a number of resonant peaks (blue-solid lines) and a smoothly varying background (black-dashed line). The black-dotted lines represent the detection efficiencies with the corresponding highest values indicated. Each vertical black-dashed line is used to collect the same resonant state but decaying to both α and p channels.$^{19}$ Ne, this work$^{19}$ Ne, literature$E_{x}$ /MeVBR $_\alpha$ $^\dagger$ (%)$\Gamma_p/\Gamma_\alpha$ $^\dagger$ $ \theta_\alpha^2$ $^\dagger$ $E_{x}$ /MeV$J^\pi$ Γ /keV $\Gamma_p/\Gamma_\alpha$ $\theta_\alpha^2$ 4.74 4.712(10)f (5/2 $^-$ )f5.44 5.424(7)f (7/2 $^+$ )f6.02 6.100(1)g (7/2 $^+$ )g6.26 6.2916(9)g (11/2 $^+$ )g6.50 6.423(3)g (3/2 $^+$ )g6.441(3)g (3/2 $^+$ )g6.89 6.853(3)g (7/2 $^-$ )g7.22 7.203(31)c (3/2 $^+$ )c35(12)c 0.67(0.23)a 7.61 63.7(4.1) 0.57(0.10) $7.3(3.5)\times10^{-3}$ 7.616(5)c (3/2 $^+$ )c21(10)c 1.2(0.4)a 7.85 54.1(2.9) 0.85(0.10) 0.073(0.027) 7.863(39)c (1/2 $^+$ )c292(107)c 1.754(14)d 0.159(54)h 8.14 45.5(2.5) 1.20(0.12) $7.8(5.7)\times 10^{-3}$ 7.974(10)c (5/2 $^-$ )c11(8)c 15.2(1.2)d 8.51 62.9(2.4) 0.59(0.06) 0.46(0.12) 8.428(2)b (13/2 $^-$ )b4(1)b 0.31(4)b 8.89 29.7(2.0) 2.37(0.23) 0.12(0.03) (8.790)b (11/2 $^{+/-}$ )b4(1)b 0.10(3)b 9.25 1.31(0.23) 9.77 9.788(13)e 10.23 $^\star$ The statistical errors are very small, but a systematic uncertainty of about 100 keV can be estimated considering mostly the energy calibration of the silicon detectors [19].$^\dagger$ Only the statistical errors (1 s.d.) are shown here. In addition, a systematic error of about$15%$ is estimated, considering the possible variation of the background function. a Calculated from$\Gamma_p$ and$\Gamma_\alpha$ results obtained in Reference [16]
b Reference [11]
c Reference [14]
d Reference [15]
e Reference [36]
f Reference [10]
g Reference [34]
h Reference [16]Table 1. Summary of the excited states of
$^{19}$ Ne, observed in the present$^6$ Li($^{16}$ O,$^{19}$ Ne$^* \to \alpha$ +$^{15}$ O)$^3$ H reaction and$^6$ Li($^{16}$ O,$^{19}$ Ne$^* \to $ p +$^{18}$ F)$^3$ H reaction. For comparison, results from previous experimental measurements are also presented.Six resonant peaks between 7.5 MeV and 9.5 MeV can be identified simultaneously in both α- and p-decay channels, as can be seen in Fig. 4 (a) and (b), respectively. For each of these states, the BW function parameters (central position and width) were maintained exactly the same for both decay channels during the above fitting procedure. The relative decay strengths (the ratios of the partial widths) can then be obtained by comparing the areas (N) of the corresponding peaks corrected by the detection efficiencies (
$ \epsilon $ ) [15]:$ \begin{align} &\Gamma_p/\Gamma_\alpha=\frac{N_p/\epsilon_p}{N_\alpha/\epsilon_\alpha}, \end{align} $
(5) The results are also listed in Table 1. In addition to the statistical errors (1 s.d.) listed in the table, a systematic uncertainty of about
$ 15% $ can be estimated, which is most likely due to the uncertainties in background modeling [20, 32] as adopted in Fig. 4.The α branch ratios (BR
$ _\alpha $ ) and the α-decay partial width$ \Gamma_\alpha $ (not listed in Table 1) above 7.6 MeV are all deduced from the measured$ \Gamma_p/\Gamma_\alpha $ ratios and the referenced total widths by neglecting the contribution of γ decay [33, 34]. Only the statistical errors are presented in Table 1. Based on the obtained$ \Gamma_\alpha $ , the dimensionless reduced α width$ \theta^2_\alpha $ can be deduced according to$ \Gamma_\alpha = 2P_L(a)\gamma_W^2(a)\theta_\alpha^2(a) $ , where$ P_L $ is the Coulomb penetration factor,$ \gamma_W^2 $ is the Wigner limit value, and a is the channel radius [23, 31, 35].$ \theta^2_\alpha $ is usually regarded as the probability of finding the α-structure in the nucleus. The obtained$ \theta^2_\alpha $ values are also listed in Table 1.It should be noted that, for the α-decay channel, there may be an accompanying background reaction-decay channel
$ ^6 $ Li($ ^{16} $ O,$ ^{15} $ O)$ ^7 $ Li$ \to \alpha \; + \; ^3 $ H, which has the same final mass combination as the$ ^{19} $ Ne-decay channel and thus cannot be excluded by the Q-value selection. However, this background channel tends to emit$ ^3 $ H- and α-particles to the same side of the beam line, whereas for the targeted physics channel, namely$ ^6 $ Li($ ^{16} $ O,$ ^{19}{\rm Ne} \to $ $ \alpha \; + \; ^{15} $ O)$ ^3 $ H, those two light particles would move to either side of the beam line. By applying the latter as a requirement, the background channel can be effectively eliminated. This has been proven by a Dalitz-plot analysis which allows to check the possible existence of the events with$ ^7 $ Li$ \to \alpha \; + \; ^3 $ H decay. This has also been checked for the p-decay channel.
New measurement of 4He and proton decays from resonant states in 19Ne
- Received Date: 2023-01-10
- Available Online: 2023-11-15
Abstract: A