Small Angle Scattering Station of BEPC Synchrotron Radiation Laboratory
- Received Date: 1900-01-01
- Accepted Date: 1900-01-01
- Available Online: 1995-03-05
Abstract: The small angle x-ray scattering (SAXS) station is located at 34m from the source in the downstream of the x-ray diffraction station. The station is equipped with a SAXS camera, a detector, an alignment carriage and on-line data acquisition and control system. The SAXS camera consists of a slit-system with position precision 1μm, a remotely controlled sample holder with vertical and horizontal translations at the plane normal to photon beam and a vacuum pipe with 0.13Pa. The research for this station is directed on the static and dynamic structure, phase composition and phase transmission in various fields such as molecular biology, materialscience, polymer chemistry and metallurgy.





Abstract
HTML
Reference
Related
PDF












DownLoad: