Abstract: In a traditional cone-beam CT system, the cost of product and computation is very high. In this paper, we propose a transversely truncated cone-beam X-ray CT system with a reduced size detector positioned off-center, in which X-ray beams only cover half of the object. The reduced detector size cuts the cost and the X-ray dose of the CT system. The existing CT reconstruction algorithms are not directly applicable in this new CT system. Hence, we develop a BPF-type direct backprojection algorithm. Different from the traditional rebinning methods, our algorithm directly backprojects the pretreated projection data without rebinning. This makes the algorithm compact and computationally more efficient. Finally, some numerical simulations and practical experiments are done to validate the proposed algorithm.