MEASUREMENT OF LOW MAGNETIC FIELD DISTRIBUTION OF A β SPECTROMETER
- Received Date: 1900-01-01
- Accepted Date: 1900-01-01
- Available Online: 1986-06-05
Abstract: The low magnetic field distribution of an iron-core double-focusing low-energy β spectrometer is measured with a second harmonic magnetic flux gate magnetometer, the relative measuring precision is better than ±0.05%. The field strength of the spectrometer is about 10G. This spectrometer is used to measure the β spectrum from tritium β decay for estimating the rest mass of electron antineutrion.





Abstract
HTML
Reference
Related
PDF












DownLoad: