Investigation of the Correction for Cascade Radiation Coincidence Summing in the Measurement of Gamma Spectrometry

Get Citation
ZHOU Feng-Qun, ZHANG Yi-Min, LUO Jun-Hua, TUO Fei and KONG Xiang-Zhong. Investigation of the Correction for Cascade Radiation Coincidence Summing in the Measurement of Gamma Spectrometry[J]. Chinese Physics C, 2007, 31(5): 487-491.
ZHOU Feng-Qun, ZHANG Yi-Min, LUO Jun-Hua, TUO Fei and KONG Xiang-Zhong. Investigation of the Correction for Cascade Radiation Coincidence Summing in the Measurement of Gamma Spectrometry[J]. Chinese Physics C, 2007, 31(5): 487-491. shu
Milestone
Received: 2006-09-11
Revised: 2006-11-15
Article Metric

Article Views(3257)
PDF Downloads(851)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Investigation of the Correction for Cascade Radiation Coincidence Summing in the Measurement of Gamma Spectrometry

    Corresponding author: KONG Xiang-Zhong,
  • Department of Physics, Pingdingshan University, Pingdingshan 467000, China2 Biochemical Engineering College of Beijing Union University, Beijing 100023, China3 Department of Modern Physics, Lanzhou University, Lanzhou 730000, China

Abstract: The correction formulas of simple, more complex and universal decay schemes about gamma-rays cascade radiation coincidence summing were deduced based on basal concepts. And then using them calculated the correction coefficient of γ-rays cascade radiation coincidence summing of 133Ba and 60Co. At the same time, it was pointed out that the correction formulas given in some literatures is improper.