MONITORING THE CONCENTRATION OF WORKING GAS IN μ-COUNTER MODEL BY CHROMATOGRAPHY
- Received Date: 1900-01-01
- Accepted Date: 1900-01-01
- Available Online: 1984-04-05
Abstract: A mixing method of working gas Ar—CH4—CH2(OCH3)2 in SQS μ-counter model and a method of monitoring the concentration of CH4, CH2(OCH3)2 in working gas by chromatography are reported. Chromatographic column coated with 15% DNP and thermal conductivity detector are used in chromatography. Using gas sampling valve, sample of working gas in SQS μ-counter model is injected, then the chromatogram peak heights are measured, The concentration of CH4 and CH2(OCH3)2 are obtained by means of comparison with standard curves.