THE MONITOR SYSTEM OF COUNTERS OF MARK-J DETECTOR
- Received Date: 1981-06-08
- Accepted Date: 1900-01-01
- Available Online: 1982-04-05
Abstract: A xenon lamp-fibre glass apparatus as a monitor system of counters of Mark-Jdetector has been built. There are 208 channels of fibre to photomultipliers of counters.The calibration of pulse amplitude of counters and the relation of time between coun-ters has been given. Due to the reference tube system, the stability of the apparatus has been improvedand the accuracy of measurement has approached to ±1%. After long time operation, we found that the apparatus are stable and reliabledesirable.