A discussion about cosmic ray mass test method of MRPC
- Received Date: 2005-09-21
- Accepted Date: 2005-11-28
- Available Online: 2006-07-05
Abstract: From the test results of 33 Multi-gap Resistive Plate Chamber (MRPC) modules, we studied the correlation between the average signal amplitude and the time resolution. The average amplitude depends on the gas multiplication operation mode which determines the time resolution. The fluctuation of time resolution measurements with the number of events collected by the chambers is also discussed. This analysis will help to set the control limits to select the chambers with required time resolution in mass production.