×
近期发现有不法分子冒充我刊与作者联系,借此进行欺诈等不法行为,请广大作者加以鉴别,如遇诈骗行为,请第一时间与我刊编辑部联系确认(《中国物理C》(英文)编辑部电话:010-88235947,010-88236950),并作报警处理。
本刊再次郑重声明:
(1)本刊官方网址为cpc.ihep.ac.cn和https://iopscience.iop.org/journal/1674-1137
(2)本刊采编系统作者中心是投稿的唯一路径,该系统为ScholarOne远程稿件采编系统,仅在本刊投稿网网址(https://mc03.manuscriptcentral.com/cpc)设有登录入口。本刊不接受其他方式的投稿,如打印稿投稿、E-mail信箱投稿等,若以此种方式接收投稿均为假冒。
(3)所有投稿均需经过严格的同行评议、编辑加工后方可发表,本刊不存在所谓的“编辑部内部征稿”。如果有人以“编辑部内部人员”名义帮助作者发稿,并收取发表费用,均为假冒。
                  
《中国物理C》(英文)编辑部
2024年10月30日

Capability Verification of the Beam Delivery System in the Superficially-Placed Tumor Therapy Terminal at HIRFL

  • The passive beam delivery system in the superficially-placed tumor therapy terminal at Heavy Ion Research Facility in Lanzhou (HIRFL), which includes two orthogonal dipole magnets as scanning system, a motor-driven energy degrader as range-shifter, series of ridge filters as range modulator and a multileaf collimator, is introduced in detail. The capacities of its important components and the whole system have been verified experimentally. The tests of the ridge filter for extending Bragg peak and the range shifter for energy adjustment show both work well. To examine the passive beam delivery system, a beam shaping experiment were carried out, simulating a three-dimensional (3D) conformal irradiation to a tumor. The encouraging experimental result confirms that 3D layer-stacking conformal irradiation can be performed by means of the passive system. The validation of the beam delivery system establishes a substantial basis for upcoming clinical trial for superficially-placed tumors with heavy ions in the therapy terminal at HIRFL.
  • 加载中
  • [1] . Kraft G. Strahlenlher Onkol, 1990, 166: 1-132. Hishikawa Y, Kagawa K et al. Ther. Radiol. Oncol, 2002,14: 73-773. LI Qiang, WEI Zeng-Quan et al. Nuclear Physics Review,2001, 18(2): 109-115(in Chinese)(李强,卫增泉等.原子核物理评论,2001,18(2): 109-115)4. Tatsuaki Kanai. Proceedings of NIRS International Semi-nar on the Application of Heavy Ion Acceleration to Radi-ation Therapy of Cancer. NIRS, Chiba, Japan, 1994. 15. Kraft G. Progresses in Particle and Nuclear Physics, 2000,45: 473-5446. ZHU Kun, ZHANG Jin-Quan et al. Nuclear Physics Re-view, 2003, 20(3): 197-200(in Chinese)(朱昆,张金泉等.原子核物理评论,2003, 20(3): 197-200)7. CHU W T, Ludewigt B A, Renner T R. Rev. Sci. Instrum.,1993, 64(8): 2088-20898. DAI Zhong-Ying, LI Qiang et al. HEP NP, 2006, 30(9):920-9249. Scholz M. Bull Cancer/Radiother, Elsevier, Paris, 1996,83(1): 50-5410. Urie M, Goitein M, Wagner M. Phys. Med. Biol., 1984, 29:55311. CHU W, Ludewigt B A, Renner T. Rev, Sci. Instruments,1993, 63: 205512. The Proton Medical Research Center. University ofTsukuba, online: www.pmrc.tsukuba.ac.jp13. Gottschalk B, Koehler A M, Wagner M. Upstream Scatter-ing Modulation in Proton Therapy Beams. PSI-Report 69.Proc. of the Int. Heavy Particle Therapy Workshop. PSI,Villigen: Switzerland, 198914. Th. Haberer. Treatment Delivery Principles for Ion Ther-apy Final Version for ENLIGHT WP 3.1 GSI. Darmstadt,Germany, 200315. Endo M, Koyama-Ito H, Minohara S et al. The Radiol.Oncol., 1996, 8: 231-23816. XIE Hong-Mei, WEI Zeng-Quan, LI Wen-Jian. NuclearTechniques, 1997, 20(4): 230-234(in Chinese)(颉红梅,卫增泉,李文建.核技术,1997, 20(4): 230-234)17. Carturight G. ShirkE. Nucl. Instrum. Methods, 1978, 153:457-46018. LI Qiang. Mutagenic Consequences of the Space Environ-ment, Cospar Colloquium. Xi'an, China, 2006. 51-52
  • 加载中

Get Citation
DAI Zhong-Ying, LI Qiang, XIAO Guo-Qing, JIN Xiao-Dong and YAN Zheng. Capability Verification of the Beam Delivery System in the Superficially-Placed Tumor Therapy Terminal at HIRFL[J]. Chinese Physics C, 2007, 31(7): 655-659.
DAI Zhong-Ying, LI Qiang, XIAO Guo-Qing, JIN Xiao-Dong and YAN Zheng. Capability Verification of the Beam Delivery System in the Superficially-Placed Tumor Therapy Terminal at HIRFL[J]. Chinese Physics C, 2007, 31(7): 655-659. shu
Milestone
Received: 2006-10-16
Revised: 2006-11-14
Article Metric

Article Views(4352)
PDF Downloads(669)
Cited by(0)
Policy on re-use
To reuse of subscription content published by CPC, the users need to request permission from CPC, unless the content was published under an Open Access license which automatically permits that type of reuse.
通讯作者: 陈斌, bchen63@163.com
  • 1. 

    沈阳化工大学材料科学与工程学院 沈阳 110142

  1. 本站搜索
  2. 百度学术搜索
  3. 万方数据库搜索
  4. CNKI搜索

Email This Article

Title:
Email:

Capability Verification of the Beam Delivery System in the Superficially-Placed Tumor Therapy Terminal at HIRFL

    Corresponding author: DAI Zhong-Ying,
  • Institute of Modern Physics, Chinese Academy of Sciences, Lanzhou 730000, China2 Graduate University of Chinese Academy of Sciences, Beijing 100049, China

Abstract: The passive beam delivery system in the superficially-placed tumor therapy terminal at Heavy Ion Research Facility in Lanzhou (HIRFL), which includes two orthogonal dipole magnets as scanning system, a motor-driven energy degrader as range-shifter, series of ridge filters as range modulator and a multileaf collimator, is introduced in detail. The capacities of its important components and the whole system have been verified experimentally. The tests of the ridge filter for extending Bragg peak and the range shifter for energy adjustment show both work well. To examine the passive beam delivery system, a beam shaping experiment were carried out, simulating a three-dimensional (3D) conformal irradiation to a tumor. The encouraging experimental result confirms that 3D layer-stacking conformal irradiation can be performed by means of the passive system. The validation of the beam delivery system establishes a substantial basis for upcoming clinical trial for superficially-placed tumors with heavy ions in the therapy terminal at HIRFL.

    HTML

Reference (1)

目录

/

DownLoad:  Full-Size Img  PowerPoint
Return
Return