Abstract:
A readout system for X-ray CCDs based on an improved architecture is presented; by optimizing several critical circuit blocks along the analog signal chain, the conflict between the readout speed and readout noise is greatly alleviated. Using CCD47-10 as its target CCD, the readout system has achieved 8.6e- readout noise and 142 eV FWHM at 5.9 keV Mn m Kα under a pixel rate of 80 kHz. Also its performance of imaging has been investigated.
-
-
References
|
[1]
|
Short A T, Keay A, Turner M J L. Proc. of SPIE, 1998, 3445: 13-27[2] Matsuura D. Development of ASICs for Multi-readout X-ray CCDs. PH.D Dissertation, Osaka Univ. Japan, 2009[3] Miyata E, Natsukari C, Akutsu D et al. Nuclear Instruments and Methods in Physics Research A, 2001, 459: 157-164[4] e2v Inc., CCD47-10 AIMO Back Illuminated Compact Pack High Performance CCD Sensor, 2006[5] LU Bo, CUI Wei-Wei, WANG Yu-Sa et al. Design of a Low Noise High Energy Resolution X-ray CCD Readout System. Nuclear Electronics Detection Technology, 2012, to be published[6] Analog Device Inc., Datasheet, Rec.C, 1999[7] Texas Instruments Inc., Datasheet, 2009[8] Razavi B. Design of Analog CMOS Integrated Circuits. NewYork: McGraw-Hill, 2000 |
-
-
Access
-
-