Abstract:
The performance of a small–cell drift chamber operated with two helium–based gas mixtures, he/c_3h_8(60/40) and he/ch_4(60/40), was studied using cosmic rays, respectively. the gas mixture ar/co_2/ch_4(89/10/1) was also tested for comparison. the spatial resolution and de/dx resolution of the helium–based gases are better than those of ar/co_2/ch_4(89/10/1). a good spatial resolution of 110μm and a de/dx resolution of 6%–7% with 30–40 samples were obtained with he/c_3h_8(60/40).
-
-
References
|
[1]
|
. Uno S et al. Nucl. Instrum. and Methods, 1993, A330: 552. Playfer S M et al. Nucl . Instrum. and Methods, 1992, A315: 4943. Cindro V et al. Nucl. Instrum. and Methods, 1991, A309: 4114. Bernardini P et al . Nucl. Instrum. and Methods, 1995, A355: 4285. CHEN Chang et al. High Energy Physics and Nuclear Physi cs, 1998,22: 787-793(in Chinese)(陈昌等高能物理与核物理,1998,22:787-793)6. Hirano H et al. Nucl. Inst rum. and Methods, 2000, A455: 2947. Pet erson D et al . Nucl. Instrum. and Methods, 2002, A478: 1428. Sciolla G et al . Nucl. Instrum. and Methods, 1998, A419: 3109. CHEN Jun et al . Nuclear Electronics Detection Technology, 2004,23: 159(in Chinese)(陈君等.核电子学与探测技术,2004,23:159)10. Jeanne D et al. Nucl. Instrum. and Methods, 1973, A111: 28711. Landau L. J. Phys. (USSR) , 1994, 8: 20112. Lehraus I et al. Nucl . Instrum. and Methods, 1982, A200: 199
|
-
-